Can I travel with the S8 Series II?
Your S8 Series II has an internal power supply that enables it to operate in other
countries. It will operate on power supplies of 100–240V and 50/60Hz. No special
adjustment is necessary, but you will require an approved power cord for the
region in which you are using the device.
Please consult the medical services department of your carrier if you intend to
use your S8 Series II on an aircraft. The S8 Series II complies with US FCC Part
15, Class B requirements.
Note: Do not use your S8 Series II while the aircraft is taking off or landing.
Can I use my S8 Series II if mains (AC) power is not available?
A DC-12 converter is available to connect the S8 Series II to a 12V or 24V DC
power source. Contact your clinician for details.
!
CAUTION
The device should not be connected to both AC and DC power sources
simultaneously.
Troubleshooting
If there is a problem, try the following suggestions. If the problem cannot be
solved, contact your equipment supplier or ResMed. Do not attempt to open the
device.
Problem/Possible Cause Solution
No display
Power is not connected. Ensure the power cable is connected and the power
switch (if available) is on.
Insufficient air delivered from S8 Series II
Ramp Time is in use. Wait for air pressure to build up or change ramp time.
Air filter is dirty. Replace air filter.
Air tubing is kinked or punctured. Straighten or replace tubing.
Air tubing is not connected
properly.
Check air tubing.
Mask and headgear are not
positioned correctly.
Adjust position of mask and headgear.
Cap(s) are missing from access
port on mask.
Replace cap(s).
Pressure required for treatment
may have changed.
See your clinician to adjust the pressure.