Exxxx Call Service (where xxxx defines an error)
or
MOTOR FAULT Call Service or UPLOAD LANGUAGES Call Service
or DATA LOST Call Service
Component failure. Call service.
The following message is displayed on the LCD after you try to update settings
or copy data to the Data Card: Card Error Remove Card
Data Card is not inserted
correctly.
Check that the Data Card is inserted into the Data Card
module with
the arrow facing up, while the S8 Series II is
switched on and not delivering therapy (standby mode).
The Data Card is only partly
inserted.
Ensure that the Data Card is inserted into the Data Card
slot as far as it will go.
You may have removed the Data
Card before settings were copied
to the S8 Series II.
Reinsert the Data Card and wait for the Settings
Success Remove Card message to appear on the LCD.
The card is not a ResScan Data
Card.
Use only a ResScan Data Card.
The following message is displayed on the LCD after you try to update settings
or copy data to the Data Card: Settings Invalid Remove Card
The identification details on the
Data Card do not match the
identification details on your
device.
Contact your clinician immediately.
The following message is displayed on the LCD after you try to update settings
or copy data to the Data Card:
Settings Error Remove Card
There is a data error on the Data
Card.
Contact your clinician immediately.
The following message is NOT displayed on the LCD after you try to update
the settings using the Data Card:
Settings Success Remove Card
The settings were not updated. Contact your clinician immediately.