24
Reprocessing the mask between patients
The mask should be reprocessed when used between patients. Cleaning,
disinfection and sterilisation instructions are available from the ResMed website,
www.resmed.com/masks/sterilization. If you do not have internet access,
please contact your ResMed representative.
Technical specications
Pressure-ow curve – The mask contains passive venting to protect against
rebreathing. As a result of manufacturing variations, the vent flow rate may vary.
Vent ow rate (L/min)
0
20
40
60
80
246810 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40
Quattro Air/ Quattro Air for Her
Quattro
Quattro FX/ Quattro FX for Her
Liberty
Mask pressure (cm H
2
O)
Quattro Air/
Quattro Air
for Her
Quattro FX/
Quattro FX
for Her
Quattro Liberty
Pressure (cm H
2
O) Flow (L/Min)
3 19 — — —
4 22 22 22 22
8 32 32 32 32
Quattro Air/
Quattro Air
for Her
Quattro FX/
Quattro FX
for Her
Quattro Liberty
Pressure (cm H
2
O) Flow (L/Min)
10 37 37 37 37
12 41 41 41 41
16 48 48 48 48
20 54 54 54 54
24 60 60 60 —
28 66 66 66 —
30 69 69 69 —
32 72 — 72 —
36 77 — 77 —
40 82 — 82 —
Deadspace: Empty volume of mask to swivel. The deadspace of the mask varies
according to cushion size.
222 mL
(Large)
243 mL
(Large)
240 mL
(Large)
151 mL
(Large)*
* Empty volume of mask to elbow.
Therapy pressure (cm H
2
O)
3 - 40 4 - 30 4 - 40 4 - 20
Resistance: Drop in pressure measured (nominal) (cm H
2
O)
at 50 L/min 0.2 0.1 0.1 0.1
at 100 L/min 0.9 0.5 0.4 0.5
Inspiratory and Expiratory Resistance (cm H
2
O)
Inspiration 0.6 0.7 0.8 0.7
Expiration 1.0 0.7 0.8 0.8