ResMed Oxygen Equipment VPAP ST User Manual

Browse online or download User Manual for Dehumidifiers ResMed Oxygen Equipment VPAP ST. ResMed Oxygen Equipment VPAP ST User's Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - User Guide

Making quality of care easyRespiratory Care SolutionsVPAP™ STNONINVASIVE VENTILATORUser GuideAEnglish 368551.indb 1 4/02/2013 3:14:20 PM

Page 2 - Making quality of care easy

8 General technical specifications Power supply 90W power supply unit Input range: 100–240V, 50–60Hz, 115V, 400Hz nominal for aircraft use Typical

Page 3 - Masks and humidifiers

English 9 VPAP ST technical specifications Mode pressure ranges CPAP mode Set Pressure: 4–20 cm H2O S, ST, T and PAC modes IPAP: 4–25 cm H2O; EPAP

Page 4 - Getting started

10  Air tubing technical specifications Air tubing Material Length Inner diameter ClimateLine heated air tubing Flexible plastic and electrical

Page 5 - Cleaning and maintenance

English 11 Humidifier performance The following settings have been tested at 22°C ambient temperature: Mask pressure cm H2O RH output % Nominal sy

Page 6 - Troubleshooting

12  Displayed values Value Range Display resolution Pressure sensor at air outlet Mask pressure 4-20 cm H2O (CPAP); 2-25 cm H2O (S, ST, T, PAC, i

Page 7 - English 5

English 13 Pressure accuracy Maximum static pressure variation at 10 cm H2O according to EN ISO 17510-1:2009 Standard air tubing SlimLine air tub

Page 8

14  Direct current; Lock/unlock; China pollution control logo 1; China pollution control logo 2; European Authorised Representative; Not d

Page 9

English 15 Product Warranty period  Clip-type finger pulse sensors  CPAP and bilevel device data modules  Oximeters and CPAP and bilevel de

Page 10

16  WARNINGS  Read the entire manual before using the device.  Use the device only as directed by your physician or healthcare provider.  Use

Page 11 - English 9

English 17  Do not leave long lengths of air tubing around the top of your bed. It could twist around your head or neck while you are sleeping.

Page 12

Making quality of care easyRespiratory Care Solutions368551.indb 2 4/02/2013 3:14:20 PM

Page 13 - Pneumatic flow path

Making quality of care easyRespiratory Care Solutions368551.indb 3 4/02/2013 3:14:20 PM

Page 14 - Displayed values

Global leaders in sleep and respiratory medicine www.resmed.com Manufacturer: ResMed Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia

Page 15 - Symbols

English 1 English Indications for use The VPAP ST is indicated to provide noninvasive ventilation for patients weighing more than 13 kg or more th

Page 16 - Limited warranty

2 Setup Refer to illustration A. 1. Connect the DC plug of the power supply unit to the rear of the device. 2. Connect the power cord to the power

Page 17 - English 15

English 3 Notes:  If your clinician has enabled SmartStart your device will start automatically when you breathe into the mask and stop automati

Page 18 - WARNINGS

4 SD card An SD card has been supplied to gather therapy data from your device and provide settings updates from your clinician. When instructed to

Page 19 - CAUTIONS

English 5 Problem/Possible cause Solution EPAP may be set too low. Talk to your clinician about your settings. Device does not start when you br

Page 20

6 Problem/Possible cause Solution The patient is using a paediatric mask with a set pressure less than 8 cm H2O. Disable SmartStart/Stop. Pressure

Page 21 - EUR1 EUR3

English 7 Problem/Possible cause Solution Displays message: No tube, please check your tube is connected Flow is high because air tubing is not c